Description
As An Equipment Monitor The 3M™ Comply™ Bowie-Dick Plus Test Pack With Early Warning Sheet 00135Lf Is Designed For Testing Air Removal Efficiency Of 132-134°C (270-273°F) Dynamic-Air-Removal Steam Sterilizers.
Test Packs Will Detect Residual Air In The Sterilizer Chamber Due To Air Leaks Air Removal Failures And Inadequate Vacuum.
The Comply Bowie-Dick Test Pack 00135Lf Features An Early Warning Sheet That Is Designed For Early Detection Of Air Removal Failures Before They Appear On The Primary Test Sheet.
Complies To Ansi/Aami/Iso 11140-52007 And Are Type 2 (Category S2) Special Indicators As Categorized Byiso 11140-12014.
Early Warning Feature Of The 3M™ Comply™ Bowie-Dick Plus Test Pack 00135Lf Permits Scheduled Sterilizer Maintenance And Helps To Prevent Unexpected Failures Indicator Sheet Will Turn A Uniform Dark Brown/Black Color Except When Air Removal Failures Such As Air Leaks Occur.
Can Be Filed For Record Keeping.
Ink Is Not Made With Lead Materials 3M™ Comply™ Bowie-Dick Plus Test Pack 00135Lf Is Equivalent In Performance To The Bowie-Dick Towel Pack Described In Ansi/Aami St79.
Two Year Shelf Life.